EPC / EPR
pH & ORP Controller · 9 models
EPC
Overview
Industrial pH/ORP controllers and recorders with high input impedance IC for high reliability and safety. Simultaneous display of measurement and set values. Simple operation for better ease of use.
Features
- One-touch calibration — insert electrode into standard solution and press CAL.
- Flexible power source — supports AC 90 to 260 V.
- Functions as both pH and ORP meter via panel switch.
- Memory backup — settings and calibration values preserved when powered off.
- Wide transmission output range — 8 ranges for pH and 8 ranges for ORP.
- Automatic electrode diagnostics — asymmetry potential, sensitivity, reaction speed.
Models
EPC-96AC
EPC-96ACT
EPC-100AC
EPC-1
EPC-2
EOC-2
EPR-1S
EPR-420S
EPR-6S
Specifications
Measurement Range
EPC-96AC
pH 0–14 · ORP 1400 mV
EPC-96ACT
pH 0–14 · ORP 1400 mV
EPC-100AC
pH 0–14 · ORP 2000 mV
Reproducibility
All EPC models
pH ±0.05 · ORP ±5 mV
Transmission Output
All EPC models
DC 4–20 mA (insulated I/O)
Max. load
500 Ω
Linearity
±0.5%
Control Contact Output
Channels
2-CH independent (IC contact)
Capacity
AC 250 V, 3 A max (resistive)
Ambient Temperature & Humidity
EPC-96AC / 96ACT
−5 to 40 °C · 85% RH or less
EPC-100AC
−5 to 50 °C · 85% RH or less
Power Supply
EPC-96AC / 96ACT
AC 90–260 V · 50/60 Hz · 10 W
EPC-100AC
AC 100–115 V or 200–240 V · 50/60 Hz · 7 VA
Dimensions
EPC-96AC / 96ACT
96 × 96 × 145 mm (panel mount)
EPC-100AC
270 × 186 × 229 mm (rain-proof)
Weight
EPC-96AC / 96ACT
Approx. 650 g
EPC-100AC
Approx. 3 kg
Applications
- Industrial pH and ORP measurement and control.
- Plating plants, chemical plants, effluent treatment.
- Food processing plants, laboratories, hospitals.
- Rain-proof outdoor installations (EPC-100AC).